Allpro ScanX-12 DVM Wartungshandbuch Seite 1

Stöbern Sie online oder laden Sie Wartungshandbuch nach Multifunktionsgeräte Allpro ScanX-12 DVM herunter. Allpro ScanX-12 DVM Technical data Benutzerhandbuch

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken

Inhaltsverzeichnis

Seite 1 - Instruction Manual

Instruction ManualPart Numbers: High Resolution Unit B7350-214SHigh Contrast Unit B7350-550G

Seite 2 - TABLE OF CONTENTS

10COMPUTER SYSTEM REQUIREMENTSIMPORTANT: To operate the ScanX, it must be connected to a compliant Computer System, and the computer mu

Seite 3 - ON-LINE REGISTRATION

11COMPUTER SYSTEM REQUIREMENTSCONTROLS & INDICATORSALLPROIMAGINGREADY ERASERFLASHINGOFFALLPROIMAGINGREADY ERASERFLASHINGOFFREADY SwitchFLASHING OF

Seite 4 - SAFETY NOTICE

12TECHNICAL DATAElectrical Requirements:Supply Voltage: 100 to 240VAC +/- 10%, 50/60 Hz Supply Current: 1.2 A MaximumLine Cord: North American styl

Seite 5

13The ScanX may be located almost anywhere in the office. Follow these guidelines for optimum performance: Lighting conditions: Set up the scanner i

Seite 6

14SYSTEM SETUPNote: Authorized Imaging Software supplied by the dealer or other company, must be installed on the computer in order to operate the Sc

Seite 7 - IMPORTANT INFORMATION

15Figure 1. Typical ScanX InstallationPower SupplyComputer Connector CablePress HERE toturn ON or OFFGreen LED IndicatorPower SupplyLine CordMain

Seite 8 - SYSTEM DESCRIPTION

16Prior to performing the imaging procedure provided on the following pages, the user must be familiar with the care, handling and preparation of the

Seite 9 - UNPACKING & INSPECTION

17PLATE CARE & PREPARATIONIMPORTANT: PSPs must always be erased prior to use.Note: Use PSPs within 24 hours of last erasure. Repeat erasing proc

Seite 10 - COMPUTER SYSTEM REQUIREMENTS

18IMAGING PROCEDURES1. Orient the cassette so that the Tube side is facing down and the hinge is away from you. 2. Open the cassette and grasp

Seite 11 - CONTROLS & INDICATORS

19IMAGING PROCEDURESScanning and Erasing Plates (continued)5. At this point, all four Scanner Track indicators will turn yellow, indicating the PSP h

Seite 12 - TECHNICAL DATA

2FOREWORDAir Techniques and its ALLPRO Imaging division have prepared this document as a guide to the proper use of the ScanX® 12 Digital Imaging Sys

Seite 13 - ABBREVIATIONS

20MAINTENANCE Maintenance ProceduresThe ScanX is designed for many years of trouble-free operation. Maintenance as described herein is minimal.IMPORTA

Seite 14 - SYSTEM SETUP

21Like all precision products, the ScanX requires a certain amount of care on a regularly scheduled basis. A well-organized maintenance p

Seite 15

22TROUBLESHOOTINGTrouble Possible Cause Corrective Action1. No power/No green light on membrane switch• Not plugged in.• No power at Mains Outlet•

Seite 16 - PLATE CARE & PREPARATION

23TROUBLESHOOTINGTrouble Possible Cause Corrective Action6. Image is too dark.• PSP has been overexposed • Use imaging software to adjust bright

Seite 17

24The following lists the ordering number and description for accessory components available to maintain the ScanX to meet your professional needs. Co

Seite 19

26NOTES

Seite 21 - SCHEDULED MAINTENANCE

© Copyright Air Techniques, Inc. 2008P/N B7319 Rev. BAir Techniques and ALLPRO Imaging are leading manufacturers of fine dental, medical and veterinar

Seite 22 - TROUBLESHOOTING

3CONGRATULATIONSWARRANTYQuickly and easily register your new ScanX on-line. Just have your product model and serial numbers available. Then

Seite 23

4SAFETY NOTICEThis equipment has been designed to minimize exposure of personnel to haz-ards. While the ScanX is designed for safe operation, cert

Seite 24 - IF YOU NEED ASSISTANCE

5SAFETY NOTICEUse of Accessory Equipment.The use of ACCESSORY equipment not complying with the equivalent safety requirements of this equipm

Seite 25

6Knowledge of Warnings and Cautions.Users must exercise every precaution to ensure personnel safety, and be familiar with the warnings and cautions pr

Seite 26

7General Notes. All instructions in this manual form an integral part of the unit. They must be kept close to the unit and in readiness whe

Seite 27

8IMPORTANT INFORMATIONCorrect Usage Operation of the ScanX may only be carried out by suitably qualified personnel. The ScanX is only to be used i

Seite 28

9UNPACKING & INSPECTIONUnpackingThe ScanX is shipped in a very special shipping case covered by a double-wall corrugated over-packaging. The case

Kommentare zu diesen Handbüchern

Keine Kommentare

Scythe Accessori per computer manuali utente

Partendo da B

Cliccare su un alfabeto di seguito per far scorrere la lista completa dei modellif models starting with that letter

Modelli Tipo di Documento
Big Shuriken 2 Rev. B
Manuale Utente   Инструкция по эксплуатации Scythe Big Shuriken 2 Rev. B, 3 pagine