Seite 1 - Instruction Manual

Instruction ManualPart Numbers: B7150 and B7350ScanX-12 DVMP/N B7350ScanX DVMP/N B7150

Seite 2 - FOREWORD

10COMPUTER SYSTEM REQUIREMENTSIMPORTANT: To operate the ScanX, it must be connected to a compliant Computer System, and the computer must be loaded

Seite 3 - ON-LINE REGISTRATION

11COMPUTER SYSTEM REQUIREMENTSCONTROLS AND INDICATORSALLPROIMAGINGREADY ERASERFLASHINGOFFALLPROIMAGINGREADY ERASERFLASHINGOFFREADY SwitchFLASHING OFF

Seite 4 - SAFETY NOTICE

12TECHNICAL DATAElectrical Requirements:Supply Voltage: 100 to 240VAC +/- 10%, 50/60 Hz Supply Current: 1.2 A MaximumLine Cord: North American sty

Seite 5

13The ScanX may be located almost anywhere in the office. Follow these guidelines for optimum performance: Lighting conditions: Set up the scanner i

Seite 6

14SYSTEM SETUPNote: Authorized Imaging Software, such as Visix or VetRay, supplied by the dealer or other company, must be installed on the computer

Seite 7 - IMPORTANT INFORMATION

15Figure 1. Typical ScanX InstallationSYSTEM SETUPPower SupplyComputer Connector CableMains OutletLine CordNote:The Line Cord is the Mains disconnect

Seite 8 - SYSTEM DESCRIPTION

16Printed sideTube or Sensitive sideFigure 2. Intraoral PSP ConfigurationPrior to performing the Intraoral and Extraoral imaging procedures provi

Seite 9 - IF YOU NEED ASSISTANCE

17PLATE CARE & PREPARATIONIMPORTANT: PSPsmustalwaysbeerasedpriortouse.Note: UsePSPswithin24hoursoflasterasure.Repeaterasingproc

Seite 10 - COMPUTER SYSTEM REQUIREMENTS

18Figure 4aInsert the PSP.Figure 4bPeel off the adhesive strip.Figure 4cSeal the Barrier Envelope.Preparing Intraoral Plates for Patient UseInsert the

Seite 11 - CONTROLS AND INDICATORS

19Note: 1. The orientation letter “a”, printed on the PSP, may be used for reference as you would use the dot on an intraoral X-ray film. In additi

Seite 12 - TECHNICAL DATA

TABLE OF CONTENTS2FOREWORDALLPRO Imaging has prepared this document as a guide to the proper use of two ScanX® Digital Imaging Systems with In-Line Er

Seite 13 - ABBREVIATIONS

204. Use the user-supplied authorized Imaging Software to activate the Scanner and to select the desired image type and resolution. 5. Verify t

Seite 14 - SYSTEM SETUP

21INTRAORAL IMAGING PROCEDURESScanning and Erasing Plates (continued)Note: Up to four PSPs can be processed simultaneously. One PSP can be inserted i

Seite 15

22EXTRAORAL IMAGING PROCEDURESFigure 11.Feeding an Extra Oral Plate into the ScanX1. Orient the cassette so that the Tube side is fac-ing down and th

Seite 16 - PLATE CARE & PREPARATION

23EXTRAORAL IMAGING PROCEDURESScanning and Erasing Plates (continued)5. Observe that a red glow eminates from the scanner exit slot.6. Repeat steps

Seite 17

24MAINTENANCE Maintenance ProceduresThe ScanX is designed for many years of trouble-free operation. Maintenance as described herein is minimal.IMPORTA

Seite 18

25Like all precision products, the ScanX requires a certain amount of care on a regularly scheduled basis. A well-organized maintenance p

Seite 19 - INTRAORAL IMAGING PROCEDURES

26TROUBLESHOOTINGNote: There should be an image on the plate only if the in-line eraser mode was disabled during scanning.Trouble Possible Cause Corre

Seite 20

27TROUBLESHOOTINGTrouble Possible Cause Corrective Action6. Image is too dark.• PSP has been overexposed • Use software to adjust bright-ness. I

Seite 21

28The following lists the ordering number and description for accessory components available to maintain the ScanX to meet your professional needs. Co

Seite 22 - EXTRAORAL IMAGING PROCEDURES

29VISIX IMAGING SOFTWAREConsider VISIX, our comprehensive digital imaging software application, providing seamless integration with our ScanX P

Seite 23

3CONGRATULATIONSCongratulations on your purchase of the ScanX-12 DVM or ScanX DVM Digital Imaging System with In-Line Erase another innovati

Seite 26 - TROUBLESHOOTING

© Copyright ALLPRO Imaging, 2006 • P/N B7157, Rev. D - December 2010ALLPRO Imaging is a leading manufacturer of imaging systems for the medical, veter

Seite 27

SAFETY NOTICE4This equipment has been designed to minimize exposure of personnel to haz-ards. While the ScanX is designed for safe operation, cer

Seite 28 - ACCESSORIES

5SAFETY NOTICEUse of Accessory Equipment.The use of ACCESSORY equipment not complying with the equivalent safety requirements of this equipm

Seite 29 - VISIX IMAGING SOFTWARE

6Knowledge of Warnings and Cautions.Users must exercise every precaution to ensure personnel safety, and be familiar with the warnings and cautions pr

Seite 30

7General Notes. All instructions in this manual form an integral part of the unit. They must be kept close to the unit and in readiness whe

Seite 31

8IMPORTANT INFORMATIONCorrect Usage Operation of the ScanX may only be carried out by suitably qualified personnel. The ScanX is only to be used i

Seite 32

9UNPACKING AND INSPECTIONUnpackingThe ScanX may be shipped in a very special shipping case covered by a double-wall corrugated over-packaging. A depos